RecallDepth

Brand Name: Perfusor Space Product Name: PERFUSOR SPACE ACM-KIT Model/Catalog Number: 8713040U Product Description: Perfusor Space, Infusion Pump Kit, 1 each Component: N/A

Company
B Braun Medical Inc
Recall Initiated
June 14, 2024
Posted
July 25, 2024
Recall Number
Z-2436-2024
Quantity
930
Firm Location
Breinigsville, PA

Reason for Recall

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

Distribution

Domestic distribution nationwide. International distribution to Canada.

Lot / Code Info

Lot Code: Model No 8713040U; UDI-DI 4046963687182; All Serial Numbers

Root Cause

Device Design

Action Taken

On June 14, 2024, the firm notified customers via URGENT MEDICAL DEVICE FIELD CORRECTION NOTIFICATION letters. Customers were advised to not use the recalled syringes in Perfusor Space Syringe Pumps. All users should be informed of the correction. Customers were provided with an attachment to post in care areas as a reminder to not use the recalled syringes with the specified pumps.

More recalls by B Braun Medical Inc