Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.

Company
Seaspine Orthopedics Corporation
Recall Initiated
February 28, 2024
Posted
April 23, 2024
Recall Number
Z-1620-2024
Quantity
444 Screw Heads
Firm Location
Carlsbad, CA

Reason for Recall

Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

Distribution

US: CA, FL, CO, TX

Lot / Code Info

REF/UDI-DI/Lots: MX1-001010/10889981298647/ AW161319D, AW164195D, AW166994D, AW167309D, AW167422D, AW169350F, AW169351F. AMRDXMIS/10889981303297/ AMRDXMIS-001, AMRDXMIS-002, AMRDXMIS-003, AMRDXMIS-004, AMRDXMIS-005

Root Cause

Nonconforming Material/Component

Action Taken

On 2/26/24, the firm phoned customers and informed them of the recall and that they would receive an email outlining details of the field action and subsequent steps to be taken. On 2/28/24, recall emails were sent to customers asking customers to respond to the emails to coordinate product return and replacement. Customers with questions can contact the firm at olivia.disanto@seaspine.com.