Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.

This recall is currently active, issued April 23, 2024. It was issued by Seaspine Orthopedics Corporation.

Company
Seaspine Orthopedics Corporation
Recall Initiated
February 28, 2024
Posted
April 23, 2024
Recall Number
Z-1620-2024
Quantity
444 Screw Heads
Firm Location
Carlsbad, CA
Official Source
View on FDA website ↗

Reason for Recall

Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

Distribution

US: CA, FL, CO, TX

Lot / Code Info

REF/UDI-DI/Lots: MX1-001010/10889981298647/ AW161319D, AW164195D, AW166994D, AW167309D, AW167422D, AW169350F, AW169351F. AMRDXMIS/10889981303297/ AMRDXMIS-001, AMRDXMIS-002, AMRDXMIS-003, AMRDXMIS-004, AMRDXMIS-005

Root Cause

Nonconforming Material/Component

Action Taken

On 2/26/24, the firm phoned customers and informed them of the recall and that they would receive an email outlining details of the field action and subsequent steps to be taken. On 2/28/24, recall emails were sent to customers asking customers to respond to the emails to coordinate product return and replacement. Customers with questions can contact the firm at olivia.disanto@seaspine.com.