Brand Name: Meridian Angled Awl Fixed Product Name: Angled Awl Fixed Model/Catalog Number: RA2-003015 Product Description: Surgical Awl Component: No

Company
Seaspine Orthopedics Corporation
Recall Initiated
June 4, 2024
Posted
July 15, 2024
Recall Number
Z-2357-2024
Quantity
100 units
Firm Location
Carlsbad, CA

Reason for Recall

Insufficient weld around the cap component of the instrument is insufficient to withstand impaction forces.

Distribution

The product was distributed to the following US states: Nevada, Colorado, Texas, Washington, Indiana, Georgia, Michigan, Florida, Pennsylvania, Illinois, Tennessee, Virginia, Missouri, South Dakota, California, Montana, Minnesota, Arizona

Lot / Code Info

Lot Code: UDI 10889981216368 Lot Numbers TT0190H, TT0191H

Root Cause

Process design

Action Taken

On June 4, 2024, SeaSpine Orthopedics issued a "Urgent Medical Device Recall" Notification via E-Mail to affected consignees. SeaSpine Orthopedics ask consignees to take the following actions: We request that you return the affected part to SeaSpine s Carlsbad facility for evaluation and potential rework. To support your upcoming cases, a replacement Vu aPod Prime NM Angled Awl will be sent (PN: 33-40-1028). Additionally, please continue to utilize the loaner pool. Please use the attached label to return the inventory in your possession at your earliest convenience. Once the inventory has been returned, please acknowledge via DocuSign.