Centurion Clamp kits labeled as: a) STERILE BACKHAUS TOWEL CLAMP (BTC173), Product Code 66520; b) STERILE TOWEL CLAMP, Product Code A547ST; c) ST TOWEL CLAMPS NON PERFORATING (P/S), Product Code I68080; d) STERILE TOWEL CLAMPS 5 1/2, Product Code I68085; e) STERILE TOWEL CLAMPS 5 1/2", Product Code I68095; f) ZIPSER PENIS CLAMP STERILE (ZC110), Product Code ZC110ST
This recall is currently active, issued March 29, 2024. It was issued by Medline Industries, Lp - Northfield.
- Recall Initiated
- February 23, 2024
- Posted
- March 29, 2024
- Recall Number
- Z-1440-2024
- Quantity
- 12980 units
- Firm Location
- Northfield, IL
- Official Source
- View on FDA website ↗
Reason for Recall
Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.
Distribution
Worldwide distribution - US Nationwide and the countries of Panama, Canada.
Lot / Code Info
a) 66520, UDI/DI 60653160040960 (case) 00653160040968 (each), Lot Numbers: 2023111590; b) A547ST, UDI/DI 10653160113683 (case) 00653160113686 (each), Lot Numbers: 2023020990, 2023033090, 2023060790, 2023092990; c) I68080, UDI/DI 20653160232602 (case) 00653160232608 (each), Lot Numbers: 2023072790; d) I68085, UDI/DI 30653160232616 (case) 00653160232615 (each), Lot Numbers: 2023032290, 2023071790; e) I68095, UDI/DI 20653160232695 (case) 00653160232691 (each), Lot Numbers: 2023062690; f) ZC110ST, UDI/DI 00653160258462 (case) 10653160258469 (each), Lot Numbers: 2023022490, 2023041090, 2023052290, 2023062890
Root Cause
Process control
Action Taken
Medline Industries, LP, issued a Medical Device Recall notice to its consignees on 02/27/2024 via email and USPS First Class mail. The notice explained the problem with the devices, risk, and requested that the products be destroyed at the consignee. Distributors were directed to notify their customers.