Non-sterile procedural trays labeled as CRANI, 2 per case

Company
Medline Industries, Lp - Northfield
Recall Initiated
November 17, 2022
Posted
January 27, 2023
Recall Number
Z-1029-2023
Quantity
12 cases (24 units)
Firm Location
Northfield, IL

Reason for Recall

Potential for sterile blades, within non-sterile SPT kits containing the blades, to puncture the outer foil layer causing a breach in the sterile packaging.

Distribution

US Nationwide distribution in the states of AR, MD, WA, WI.

Lot / Code Info

Item Number: DYNJ900413P, UDI/GTIN Case: 40195327031665, UDI/GTIN Each: 40195327031665, Lot Number: 22JDB689

Root Cause

Device Design

Action Taken

Medline Industries, LP notified consignees via email and first-class mail on 11/21/2022. Consignees were instructed to over-label the sterile kits with a warning label indicating Warning To Avoid Injury, Please Handle Blades with Care . For the Non-Sterile kits, they are to over label the affected kits indicating to remove the blades and replace from a different sterile supply. No product is to be returned. Customers were also instructed to complete and return the response form.