Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-10, REF G21049; b) HEMO-10-EU, REF G21346

Company
Wilson-Cook Medical Inc
Recall Initiated
May 22, 2023
Posted
June 21, 2023
Recall Number
Z-1979-2023
Quantity
14,877 units
Firm Location
Winston Salem, NC

Reason for Recall

There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue and consequent difficulty or inability to maneuver/remove the endoscope.

Distribution

US Nationwide - Worldwide Distribution

Lot / Code Info

UDI/DI : a) 00827002210493, b) 00827002213463; All unexpired lot numbers

Root Cause

Under Investigation by firm

Action Taken

The firm notified their consignees by letter and email beginning on 05/22/2023. The notice explained the issue, risk, and requested the following: Complete and return the Acknowledgement Form, ensure the notification is shared with appropriate personnel, down to the user level or customers if devices were further distributed, report any adverse events.