Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-3X6 G48278; Fusion Lithotripsy Extraction Baskets are intended to be used for endoscopic removal of biliary stones and foreign bodies.
- Company
- Wilson-Cook Medical Inc
- Recall Initiated
- June 22, 2023
- Posted
- July 28, 2023
- Recall Number
- Z-2253-2023
- Quantity
- 402 units
- Firm Location
- Winston Salem, NC
Reason for Recall
Non-conforming product may appear in the field as inability to advance or retract the basket, or the drive wire breaking or separating from handle (with or without mechanical lithotripsy).
Distribution
Worldwide distribution.
Lot / Code Info
UDI/DI 00827002482784, 10827002482781, Lot Numbers: W4631139, W4702268, W4631139, W4631139, W4693489, W4693489, W4576227, W4702262, W4693489, W4616511, W4693489, W4612934, W4609498, W4693489, W4609498, W4576227, W4631537, W4644272, W4693489, W4702262, W4576227, W4616511, W4612934, W4576227, W4631537, W4644272, W4631537, W4616511, W4702262, W4576227, W4612934, W4576227, W4612934, W4612934, W4612934, W4693489, W4631537, W4644272, W4609498, W4609498, W4702262, W4576227, W4612934, W4576227, W4631537, W4697764, W4576227, W4616511, W4631537, W4697764, W4644272, W4697764, W4702262, W4616511, W4697764, W4702262, W4660177, W4616520, W4610273, W4616520, W4616520, W4635963, W4622445, W4662350, W4645764, W4616520, W4625161, W4701899, W4622445, W4647285, W4647285, W4647285, W4635963, W4645764, W4663702, W4701899, W4645764, W4554892, W4663702, W4701899, W4702270, W4645764, W4616520, W4622445, W4645764, W4660177, W4645764, W4701897, W4702270, W4662350, W4663702, W4662350, W4610273, W4554892, W4664043, W4589622, W4590893, W4590893, W4590893, W4594583, W4594583, W4594583, W4594583, W4594583, W4619627, W4619627, W4619627, W4619651, W4625160, W4625160, W4625160, W4631933, W4647697, W4647701, W4651755, W4651755, W4651765, W4651765, W4651765, W4664574, W4664574, W4660177, W4645764, W4622445, W4610273, W4610273, W4616520, W4625161, W4701899, W4554892, W4554892, W4622445, W4664043, W4701899, W4625161, W4647285, W4554892, W4647285, W4622445, W4702270, W4554892, W4554892, W4554892, W4635963, W4701897, W4616520, W4631139, W4631139, W4631139, W4693489, W4631139, W4644272, W4697073, W4697252, W4631535, W4616511, W4621105, W4609498, W4616511, W4609498, W4554892, W4610273, W4664042, W4693489, W4625161, W4662350, W4610273, W4635963, W4635963, W4647285, W4664043, W4625161, W4660177, W4663702, W4664043, W4701897, W4701899, W4702270, W4662350, W4662350, W4663702, W4664043, W4701897, W4609498, W4664042, W4663702, W4697764, W4664042, W4645764, W4635963, W4702262
Root Cause
Process control
Action Taken
Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice to their consignees on 07/14/2023 by email or priority mail. The notice explained the issue, the risk, and requested the consignee cease use and return the affected product.