3M Red Dot Monitoring Electrode with Foam Tape - Product Usage: intended to be used by healthcare professionals for ECG monitoring. 50 electrodes per pouch; 20 pouches per shipper box (1000 electrodes total).

Company
3M Company Health Care Business 3M Center
Recall Initiated
September 15, 2020
Terminated
February 1, 2022
Recall Number
Z-0423-2021
Quantity
720,000
Firm Location
Saint Paul, MN

Reason for Recall

Corrosion could cause performance failures in the product.

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Europe (Austria, Belgium, Switzerland, Germany, Spain, Finland, France, Greece, Ireland, Italy, Netherlands, Norway, Portugal, Sweden) and Worldwide (China, Taiwan, Singapore, Indonesia, Japan, Mexico, Peru, Costa Rica, Canada).

Lot / Code Info

Cat. No. 2228 (SKU: 70-2011-8514-0, SAP # 7100212952). GTIN: 10707387789524 (pouch) and 50707387789522 (shipper)  Lots: 202204MA (Expiration 2020-04-27)

Root Cause

Process control

Action Taken

The recalling firm issued customer letters to consignees on 10/7-8/2020. The recalling firm is requesting a returned acknowledgement form from the consignee. Distributors will also be sent a template customer letter and customer acknowledgement so that they may notify their customers. The recalling firm will initiate multiple attempts to follow up with non responders. Customers and distributors are requested to dispose of any products still in their possession.