RecallDepth

IntelliVue MX700 patient monitor, model no. 865241

This recall has been terminated (originally issued June 30, 2020).

Company
Philips North America, Llc
Recall Initiated
June 30, 2020
Terminated
January 25, 2024
Recall Number
Z-2697-2020
Quantity
28
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.

Distribution

Domestic distribution to CA, HI, MA, NY, and UT. International distribution to Costa Rica, Ireland, Lebanon, and Poland.

Lot / Code Info

Serial numbers DE58562415 DE58562344  DE58562353

Root Cause

Process control

Action Taken

On June 30, 2020, the firm sent Urgent Medical Device Correction letters to affected customers. Customers were instructed to always ensure that the monitor is grounded during operation with mains power. Affected products should be removed from use until the power supply of the monitor has been replaced. Philips Customer Service will contact customers when a replacement power supply is ready for installation. Philips expects all replacements to be complete within six months. If you need any further information or support concerning this issue, please contact your local Philips representative or Philips Customer Support at 1-800- 722-9377.

More recalls by Philips North America, Llc

View all recalls by this company →