NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)07332747010619

Company
Nobel Biocare Usa Llc
Recall Initiated
July 19, 2019
Terminated
October 27, 2020
Recall Number
Z-2207-2020
Quantity
221 implants
Firm Location
Yorba Linda, CA

Reason for Recall

The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.

Distribution

US: AL, AR, AZ, CA, CO, CT, FL, HI, ID, IL, IN, KY, MA, MS, MI, MO, NC, NH, NV, NY, OK, PA, SC, SD, TX, UT, WA, WI OUS: None

Lot / Code Info

Lot number: 13080792

Root Cause

Labeling mix-ups

Action Taken

Customers who received product/affected lot were to be contacted via Urgent Medical Device Recall Letter dated July 18, 2019 via USPS on July 19, 2019. In addition to informing customers about the recall, the recall notification asked customer to take the following actions: 1. Identify affected devices in your stock by confirming the information on the side label: NobelActive Internal NP 3.5x15mm, article 34128, lot 13080792. 2. Confirm the information on the top label as shown in figure 1 above. 3. Complete attached Customer Acknowledgment Form and return it to Nobel Biocare, via email to us.complaints@nobelbiocare.com or fax 714-282-5073, within 5 days of receipt of this letter. 4. Return incorrectly labeled products (top label 3.5x11.5) to: Nobel Biocare Attn: Returns PFA1917 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 . 5. Correctly labeled products (top label 3.5x15) do not need to be returned. 6. If you require any further information or support, please contact your customer support representative at 1-800-322- 5001 x.1527.

More recalls by Nobel Biocare Usa Llc

View all recalls by this company →