HeartStart XL+ Defibrillator/Monitor, Model 861290

This recall has been terminated (originally issued November 19, 2019).

Company
Philips North America, Llc
Recall Initiated
October 16, 2019
Posted
November 19, 2019
Terminated
November 29, 2023
Recall Number
Z-0498-2020
Quantity
23,749 Units
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

The rotary therapy selector switch may fail, resulting in unexpected device behavior. These behaviors include: The device may not turn on; The device may not perform the selected function; The device may deliver a shock with an energy level different from the setting selected by the user.

Distribution

US Nationwide

Lot / Code Info

All units manufactured prior to 1 May 2017.

Root Cause

Process control

Action Taken

On October 16, 2019, Philips Healthcare issued Urgent Medical Device Correction notices to U.S. customers via certified mail. Philips Markets Organizations are responsible for distributing the letters outside of the U.S. Actions to be taken by the customer: 1) The device is safe to use and can remain in service if it does not exhibit any of these behaviors described in this Notice. 2) Continue to perform Shift Checks and Operational checks as recommended in the Instructions for Use (IFU) as this reduces the risk of a failure during use. 3) If you identify a device that exhibit any of these behaviors; please remove it from service and contact Philips to request service. Philips will contact you to arrange for repair of your unit once parts are available. Philips will install a replacement switch in affected devices at no charge to the customer. If you need further information or support concerning this notification, please contact your local Philips representative.

More recalls by Philips North America, Llc

View all recalls by this company →