RecallDepth

SwishTapered Implant

Company
Implant Direct Sybron Manufacturing, Llc
Recall Initiated
December 12, 2016
Posted
January 28, 2017
Terminated
May 22, 2017
Recall Number
Z-1112-2017
Quantity
17
Firm Location
Westlake Village, CA

Reason for Recall

The extender product code (9037-03) that is packaged with the SwishTapered (Part Number 933308; Lot Number 73241) Implant was packaged in the incorrect position. This discrepancy may lead to the possibility of extender not seating properly on the implant, which may lead to micro-leakage and a possible infection to the patient.

Distribution

U.S. distribution to the following; NM, TX, CA, OR, OH, ME. Foreign distribution to the following; Netherlands.

Lot / Code Info

933308 Lot # 73241

Root Cause

Packaging process control

Action Taken

Customer notification letters were sent on 12/12/16. Customers were requested to review inventory for the affected product. Complete and return the Acknowledgement and Recall Return Form within 48 hours. If customers are an authorized distributor the firm requests that they identify those customers that may have been shipped the affected product lot and contact those customers to inform them of the issue within 48 hours of receipt of the notification in order to provide the customers with replacement product. If customers have any of the affected product the firm will send a replacement part. Any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425.

More recalls by Implant Direct Sybron Manufacturing, Llc