Thio Med w/Dex, Hem, Vit K (7ml) 20/PK Product Usage; Recommended for use in qualitative procedures as a general purpose media for the cultivation of aerobes and anaerobes

Company
Remel Inc
Recall Initiated
April 17, 2019
Terminated
August 21, 2020
Recall Number
Z-1379-2019
Quantity
133 units
Firm Location
Lenexa, KS

Reason for Recall

Potential for products to fail performance testing per the Instructions for Use (IFU) starting at 57 days after date of manufacturing. Expiration date is shortened from 7 months to 56 days.

Distribution

Worldwide Distribution -US Nationwide distribution to AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, VA, VT, WA, WI, WV, WY, Puerto Rico. International distribution to Singapore

Lot / Code Info

Lot Number (Expiration Date): 303285 (09/27/2018), 309151 (10/10/2018), 353575 (01/02/2019).

Root Cause

Under Investigation by firm

Action Taken

Thermo Fisher Scientifica/ Remel notified consignees on about 04/17/2019 via " URGENT MEDICAL DEVICE FIELD ACTION RECALL" letter. Instructions included to locate all affected in inventory, destroy any affected product identified, contact Customer Services or local distributor to arrange for replacements, notify all customers if product was further distributed, and complete and return the Field Action Response form. Customers were also instructed that the requirement for review of reported test results using product beyond the updated expiry date should be determined by the appropriate technical expert at each customer location. Questions or concerns can be directed to Technical Services Department at 800-255-6730 (US) or 913-888-0939 (International).