COOK MEDICAL Captura SERRATED BIOPSY FORCEPS WITH SPIKE,2.4 mm, 230 cm, with Spike, DBF-2.4SN-230SP-20-S
- Company
- Wilson-Cook Medical Inc
- Recall Initiated
- December 18, 2018
- Terminated
- December 14, 2020
- Recall Number
- Z-0774-2019
- Quantity
- 8 units
- Firm Location
- Winston Salem, NC
Reason for Recall
The product is being recalled because Captura Serrated Forceps with Spike (DBF-2.4SN-230SP-20-S) were mislabeled as Captura Serrated Large Forcep-No Spike (DBF-2.4SL-230-20-S) and Captura Serrated Large Forcep-No Spike (DBF-2.4SL-230-20-S) were mislabeled as Captura Serrated Forceps with Spike (DBF-2.4SN-230SP-20-S).
Distribution
France, Ireland, Japan
Lot / Code Info
Lot Number W4035256
Root Cause
Error in labeling
Action Taken
The firm initiated the recall by letter in Europe on 18-December-2018 and 20-December-2018. The notices were sent in Japan today, 21-December-2018. The letter requested recall to the user level and return of the affected product.
More recalls by Wilson-Cook Medical Inc
Teslatome Bipolar Sphincterotomes
REF: TESLA-B2535-260
Model Number: G60784
UDI-DI code: 00827002607...
Feb 13, 2026