Microtek Franz Angiography Drape with Radiation Shield, Item # RBA202, and Microtek Angiography Drape with Radiation Shield and Pouches, Item # RBA203. Sterile EO. Microtek Medical, Inc., Columbus, MS 39702.

This recall has been terminated (originally issued July 14, 2016).

Company
Ecolab Inc
Recall Initiated
July 14, 2016
Terminated
November 2, 2016
Recall Number
Z-2744-2016
Quantity
6750 eaches (675 cases)
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Ecolab is recalling 2 models of Angiography Drapes with Radiation Shields because the sterile packaging may contain small holes in the film of the pouch.

Distribution

US: OH, NY, CA, MN.

Lot / Code Info

RBA202 Lot numbers: D111574, D111654, D111724, D111864, D112024, D112094, D112224, D120164, D120454, D120544, D121144, D121314, D121674, D121944, D122204, D122554, D122854, D123324, D130354, D130454, D130914, D131704, D131994, D132464, D132974, D140834, D141204, D141774, D142244, D150404, D151334, D151464, D152304, D152364, D152454, D153084, D153204, D153494, D160544.   RBA203 Lot numbers: D122194, D122354

Root Cause

Packaging process control

Action Taken

Consignees were sent on 7/14/2016 an Ecolab "Urgent Medical Device Recall" letter dated July 14, 2016. The letter described the problem and the product involved in the recall. Advised consignees to immediately quarantine the product and to notify their customers. For questions contact Customer Service at 1-800-824-3027.