RecallDepth

Slush + Warmer Disc Drape; Model ORS-320 Slush + Warmer Disc Drapes - Product Usage: single-use equipment cover intended for use during various surgeries where warm irrigation, slush and/or cold solution is required.

This recall has been terminated (originally issued January 11, 2021).

Company
Ecolab Inc
Recall Initiated
January 11, 2021
Terminated
January 9, 2023
Recall Number
Z-1141-2021
Quantity
132 cases (3168 drapes)
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Slits were identified in a low quantity of ORS-320 Slush + Warmer Disc Drapes from a single lot. A source for potential manufacturing error was identified and investigation is underway to prevent further recurrence. The slits may lead to a breach in the sterile barrier between fluid in the basin and the slush/warming equipment. This could pose a risk of infection to patients receiving warmed irrigation fluid or surgical slush if used during a procedure.

Distribution

Worldwide distribution - US Nationwide Distribution in the states of AL, AZ, CA, CO, FL, HI, ID, IL, IN, KS, LA, MA, MD, MI, MO, MT, NJ, NY, OH, OK, OR, PA, TN, TX, UT, WA. In the country of Japan.

Lot / Code Info

Model ORS-320, lot D192619A, 132 cases (3168 drapes)

Root Cause

Process control

Action Taken

The firm sent a recall communication letter "Urgent: MEDICAL DEVICE RECALL IMMEDIATE ATTENTION REQUIRED" to customers on 01/11/2021.