RT-5100 (Epic 5100) Refractor and RT 3100 Refractor: Product Usage: Instrument providing means of positioning spherical and cylindrical lenses, prisms and other optical devices in front of a subjects eye for purpose of determining refractive error and binocular functions.

This recall has been terminated (originally issued March 24, 2016).

Company
Nidek Inc
Recall Initiated
February 29, 2016
Posted
March 24, 2016
Terminated
September 24, 2018
Recall Number
Z-1245-2016
Quantity
2703 separate serial numbers
Firm Location
Fremont, CA
Official Source
View on FDA website ↗

Reason for Recall

There has been a reoccurrence of an issue on the RT 5100 and RT 3100 Refractor of the Near Point Chart Arm lowering spontaneously with potential for slight injury to face or nose of patient.

Distribution

US Nationwide Distribution

Lot / Code Info

serial numbers: See attached distribution list. 128 RT -3100 and 2565 RT-5100.

Root Cause

Component design/selection

Action Taken

Nidek sent a Recall letters dated February 29, 2016 to all affected site. The letter identified the affected product, problem and actions to be taken. Nidek Customer Service will be replacing the affected product with new parts to address this recall. For questions call Customer Service at 1-800-722-0219.