Digital Radiography System Model DFP-8000D/FPD,
- Company
- Toshiba American Med Sys Inc
- Recall Initiated
- November 10, 2005
- Posted
- December 23, 2005
- Terminated
- February 5, 2007
- Recall Number
- Z-0325-06
- Quantity
- 13
- Firm Location
- Tustin, CA
Reason for Recall
To correct a software problem that caused the LV images, which were acquired for 15 seconds at the end of the examination, to be lost.
Distribution
Nationwide
Lot / Code Info
Serial Numbers: B5522165 B4592121 B4602124 B4602126 W1B0542172 W1B0542174 B4582115 B4622147 B4572111 B5522161 B5512160 A4562105 B4582113
Root Cause
Other
Action Taken
An Advisory Letter was sent to the customers via certified mail. Toshiba issued a Field Modification Instruction (FMI XRA29-90828) to correct that software bug and bring the DFP-8000D into compliance. The FMI is provided to the customers at no charge.