Stryker Secure II hospital bed; model 3002.
This recall has been terminated (originally issued June 12, 2003).
- Company
- Stryker Medical
- Recall Initiated
- May 1, 2003
- Posted
- June 12, 2003
- Terminated
- June 11, 2003
- Recall Number
- Z-0868-03
- Quantity
- 6
- Firm Location
- Kalamazoo, MI
- Official Source
- View on FDA website ↗
Reason for Recall
A wire may break at the crimp and present a potential electric shock hazard.
Distribution
Pennsylvania and Kansas
Lot / Code Info
serial numbers 030415710, 030415711, 030415712, 030415713, 030415714 and 030415729.
Root Cause
Other
Action Taken
The firm sent service representatives to the receiving hospitals, beginning on 5/1/02, and the beds were corrected in place.