Stryker brand Rugged EZ-Pro R4 Ambulance Cot, model 6092.

This recall has been terminated (originally issued October 20, 2004).

Company
Stryker Medical
Recall Initiated
August 17, 2004
Posted
October 20, 2004
Terminated
October 19, 2004
Recall Number
Z-0023-05
Quantity
15
Firm Location
Kalamazoo, MI
Official Source
View on FDA website ↗

Reason for Recall

Locking pins may be installed backward, and cot may drop from highest to lowest position when the users are not prepared for the drop.

Distribution

Alabama, California, Florida, Kansas, Montana, Oklahoma, Texas and Virginia.

Lot / Code Info

Serial numbers 040639739, 040639849, 040639737, 040639738, 040639675, 040639674, 040639670, 040639671, 040639672, 040639716, 040639847, 040639848, 040639935, 040639936 and 040639937.

Root Cause

Other

Action Taken

Service representatives visited each customer, inspected the cots, and reinstalled the lock pins, if required, beginning on August 10, 2004.