Stryker brand Rugged EZ-Pro R4 Ambulance Cot, model 6092.
This recall has been terminated (originally issued October 20, 2004).
- Company
- Stryker Medical
- Recall Initiated
- August 17, 2004
- Posted
- October 20, 2004
- Terminated
- October 19, 2004
- Recall Number
- Z-0023-05
- Quantity
- 15
- Firm Location
- Kalamazoo, MI
- Official Source
- View on FDA website ↗
Reason for Recall
Locking pins may be installed backward, and cot may drop from highest to lowest position when the users are not prepared for the drop.
Distribution
Alabama, California, Florida, Kansas, Montana, Oklahoma, Texas and Virginia.
Lot / Code Info
Serial numbers 040639739, 040639849, 040639737, 040639738, 040639675, 040639674, 040639670, 040639671, 040639672, 040639716, 040639847, 040639848, 040639935, 040639936 and 040639937.
Root Cause
Other
Action Taken
Service representatives visited each customer, inspected the cots, and reinstalled the lock pins, if required, beginning on August 10, 2004.