REFLECTION(R), MICROSTABLE(R), ACETABULAR LINER, 20o OVERHANG, 28 MM ID, 54-56 OD, QTY: (1), STERILE EO, REF 71742854, Smith & Nephew, Inc. Memphis, TN 38116 USA. Orthopaedic.
This recall has been terminated (originally issued December 26, 2013).
- Company
- Smith & Nephew Inc
- Recall Initiated
- November 25, 2013
- Posted
- December 26, 2013
- Terminated
- February 1, 2017
- Recall Number
- Z-0581-2014
- Quantity
- 38 units
- Firm Location
- Memphis, TN
- Official Source
- View on FDA website ↗
Reason for Recall
One batch of the Reflection(TM) Polyethylene Acetabular Liners may be out of tolerance causing the liner not to fully seat into the shell.
Distribution
Distributed in China and Spain.
Lot / Code Info
Batch No. 12BM04836
Root Cause
Other
Action Taken
All affected sales staff was notified of problem and the recall via email on 11/25/2013. There were no US hospitals affected by this recall.