Stryker Titanium Intraline Anchor, Model Numbers 3910400050, 3910400055, 3910400060, 3910400065, Soft Eyelet Anchor, manufactured by Stryker Endoscopy, Puerto Rico

This recall has been terminated (originally issued December 14, 2010).

Company
Stryker Endoscopy
Recall Initiated
August 20, 2009
Posted
December 14, 2010
Terminated
December 14, 2010
Recall Number
Z-0618-2011
Quantity
447 units
Firm Location
San Jose, CA
Official Source
View on FDA website ↗

Reason for Recall

Potential breakage -- product may break while inserting it into the bone during surgery. This may prolong surgery, or, if undetected during surgery, result in retained fragments that may cause metal sensitization or immune response, or necessitate further intervention to retrieve the titanium fragments.

Distribution

Nationwide Distribution: to one distributor in NJ.

Lot / Code Info

Model Numbers 3910400050, 3910400055, 3910400060, 3910400065, serial numbers 09049AE2, 09050AE2, 09051AE2, 09052AE2, 09054AE4

Root Cause

Material/Component Contamination

Action Taken

Stryker issued an Urgent Device Removal notification letter dated August 21, 2009 to consignees. Consignees were asked to acknowledge receipt and distribute the letter, track reponses, and destroy any returned material and document through weekly summary and records. Stryker can be contacted at 408 754-2000 about this recall.