Heater Elements for Resuscitaire Infant Radiant Warmers model numbers RW82, RW82 VHA, WBR82, and WMRW82, 120 volt configuration only.

Company
Draeger Medical Systems, Inc.
Recall Initiated
July 29, 2009
Posted
August 26, 2009
Terminated
October 13, 2010
Recall Number
Z-1693-2009
Quantity
2915 elements
Firm Location
Telford, PA

Reason for Recall

Spark, flame and smoke: A mis-positioned crimp on heater element lead wire resulted in flame and smoke aerating from the top of the warmer assembly.

Distribution

Worldwide distribution: USA, Canada, Saudi Arabia, Bermuda, Virgin Islands, Mexico, Panama, Colombia, and Costa Rica.

Lot / Code Info

Model numbers RW82, RW82 VHA, WBR82, and WMRW82. Date codes ending in 07

Root Cause

Process control

Action Taken

Draeger Medical issued an Urgent Medical Device Recall letter via certified mail on 8/7/09. The letter explained the problem, the need to conduct an inspection of the infant warmer (instructions included), and request replacement elements.