Extron Electronics OEM, CAB, DVI-Male to DVI-Female Cable Assembly (14 inch.); Part #: 43-111-01 (Stryker Part #: 0100-224-131); Distributed by Stryker Communications, Flower Mound, TX 75028. (These cables are commonly used with the Switchpoint Infinity II)
This recall has been terminated (originally issued January 12, 2008).
- Company
- Stryker Communications
- Recall Initiated
- October 3, 2007
- Posted
- January 12, 2008
- Terminated
- June 16, 2008
- Recall Number
- Z-0302-2008
- Quantity
- 2,255 cables.
- Firm Location
- Flower Mound, TX
- Official Source
- View on FDA website ↗
Reason for Recall
Intermitten or Complete Loss of Signal: out-of-specification DVI cables causing intermitten or complete loss of signal on monitors used in surgery.
Distribution
Worldwide: USA and Australia, Belgium, Canada, France, Germany, Great Britain, Hong Kong, Japan, Korea, Mexico, Netherlands, New Zealand, Portugal, South Africa, Spain, Sweden, Switzerland, Taiwan, and The U.A.E.
Lot / Code Info
All Lots and Serial Numbers shipped between Oct. 2006 and June 2007.
Root Cause
Nonconforming Material/Component
Action Taken
Firm notified consignees of recall via letter (Urgent Device Recall) on 09/21/07. Consignees were notified that a customer service representative would visit each facility to replace out-of-spec cables with properly shielded cables at no charge.