ProSpecT Shiga Toxin E. coli (STEC) Micorplate Assay. Catalog #: 2474048 (48 well plate) and 2474096 (96 well plate). For qualitative detection of Shiga toxins (STX1 & STX 2) in aqueous extracts of fecal specimens and fecal enriched broth cultures.
This recall has been terminated (originally issued August 17, 2004).
- Company
- Remel Inc
- Recall Initiated
- July 18, 2002
- Posted
- August 17, 2004
- Terminated
- September 15, 2004
- Recall Number
- Z-1373-04
- Quantity
- 83 (15-2474048, 68-2474096)
- Firm Location
- Lenexa, KS
- Official Source
- View on FDA website ↗
Reason for Recall
A potential contamination was noted in the conjugate in the ProSpecT Shiga Toxin E. coli Microplate Assay. The particulate in the conjugate may cause the dropper tip to become plugged.
Distribution
US and OUS.
Lot / Code Info
918706 918707
Root Cause
Other
Action Taken
Remel contacted each customer with a letter, by mail, requesting a count of the affected lots to be returned. Product would be replaced upon customer request.