Deflectable Tip Catheter for cardiac mapping. Mfg Part # Catalog # EU Part # D-1078-35-S 107835S SW1078-035; D-1078-63-S D5S06AL252RT 36A-07Q; D-1078-78-S 107878S N/A; D-1097-511-S 1097511S SW1097-511; D-1097-554-S D708DL002RT 36F-32Q.

Company
Biosense Webster, Inc.
Recall Initiated
August 16, 2004
Posted
September 2, 2004
Terminated
March 31, 2005
Recall Number
Z-1437-04
Quantity
848
Firm Location
Irwindale, CA

Reason for Recall

Potential that product is incorrectly assembled with monel wire attached to the tip dome rather than copper wire.

Distribution

Nationwide and Japan, Europe, & Uruguay

Lot / Code Info

112 lots

Root Cause

Other

Action Taken

Sales Representative to provide letter to each of the 25 US customers explaining the action. Sales Representatives will remove any affected product from customers inventory. Certification records will be completed for each customer confirming the removal action. International affiliates to conduct similar actions.