GE Centricity Ultra Laboratory System Software; for recording, reporting and distribution of lab results. GE Healthcare Integrated IT Solutions, Barrington, IL 60010.

This recall has been terminated (originally issued December 31, 2008).

Company
Ge Healthcare Integrated It Solutions
Recall Initiated
September 22, 2008
Posted
December 31, 2008
Terminated
May 18, 2010
Recall Number
Z-0472-2009
Quantity
3 units
Firm Location
Barrington, IL
Official Source
View on FDA website ↗

Reason for Recall

Software computer, error: The abnormal result flag is not being sent in the HL7 messages on results that are lower than the normal range. For further information, please telephone GE Healthcare Integrated IT Solutions at 800-437-1171.

Distribution

Globally - United States including the states of New York and Texas; and the country of Canada.

Lot / Code Info

Software versions 3.3.2.27_P4 and 4.0.1_SP.x

Root Cause

Software design

Action Taken

GE Healthcare notified their Centricity Ultra Laboratory Information System customers via letter dated 9/19/08. The letter was entitled "PRODUCT SAFETY NOTIFICATION." The letter states that the software correction was installed or will be installed to correct the abnormal results flag not being sent in the HL7 on results that are lower than the normal range. The letter summarized the safety concern and the action taken. Any questions were directed to GE Technical Support at 888-778-3373 or 905-479-1249. The accounts were requested to complete the enclosed customer reply form, supplying the software version number, and fax it back to 847-277-5240.