AccuData GTS Plus AccuData GTS For Blood Glucose Monitoring and Automated Data Management, contents Accu-Chek Advantage repalcement Meter and module, cat. no. 454; Ref. no. 03144798001 contains meter model no. 777.

This recall has been terminated (originally issued August 5, 2006).

Company
Roche Diagnostics Corp.
Recall Initiated
June 22, 2006
Posted
August 5, 2006
Terminated
April 7, 2008
Recall Number
Z-1367-06
Quantity
55,189 meters (for all Recalls Z-1354-1371-06)
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

The meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.

Distribution

Worldwide, USA, Germany, Guam, North Mariana Islands, Panama and Virgin Islands.

Lot / Code Info

All units.

Root Cause

Other

Action Taken

A recall notification letter dated April 24, 2006 was sent to each customer alerting them to the fact that this error code has a dual meaning. New software will be installed in all meters when it becomes available.