Boston Scientific Flexima Biliary Stent System, 200 cm long, .035" (0.89 mm), sterile, Boston Scientific, Spencer, IN; REF 3921, UPN M00539210. System is used for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.

This recall has been terminated (originally issued September 11, 2008).

Company
Boston Scientific Corp
Recall Initiated
April 2, 2008
Posted
September 11, 2008
Terminated
September 11, 2008
Recall Number
Z-2105-2008
Quantity
80
Firm Location
Spencer, IN
Official Source
View on FDA website ↗

Reason for Recall

Rapid Exchange Biliary Stent may be mislabeled as Flexima biliary Stent, which has a different delivery system included in the package.

Distribution

Class III Recall - Worldwide Distribution --- including USA and countries of Finland, France, Germany and Great Britain.

Lot / Code Info

Lot 11374425.

Root Cause

Packaging process control

Action Taken

An Urgent Medical Device Recall-Immediate Action Required letter was sent to consignees on 4/2/08 directing them to immediately discontinue use of and segregate the identified products for return to the firm and complete the verification tracking form.