Philips Medical HeartStartXL Defibrillator/Monitor Model : M4735A

Company
Philips Medical Systems
Recall Initiated
June 11, 2007
Posted
August 7, 2007
Terminated
August 8, 2016
Recall Number
Z-1128-2007
Quantity
312 units
Firm Location
Andover, MA

Reason for Recall

The ECG signal may become unusable to assess patient condition and need for therapy.

Distribution

Worldwide, including USA, Canada, Australia, Brazil, China, Czech Republic, France, Germany, Hong Kong, India, Ireland, Italy, Japan, Korea, Kuwait, Macau, Poland, Portugal, Romania, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, UK, and Venezuela.

Lot / Code Info

Serial Numbers: US00453441 through US00453910

Root Cause

Nonconforming Material/Component

Action Taken

Philips Medical contacted accounts by letter dated June 11, 2007. Users are advised to remove product from service, if possible. If not, a back up unit should be available to use. Service reps will contact accounts to replace units.