Patient tables of CT systems with brand name: CT/e; CT/e Dual; CT/e Plus, CT/e Dual Plus; CT/e Lite; ProSpeed AI; ProSpeed AII; ProSpeed FI; ProSpeed FII; ProSpeed EII; CT HiSpeed Series.
This recall has been terminated (originally issued February 24, 2005).
- Company
- General Electric Med Systems Llc
- Recall Initiated
- October 21, 2004
- Posted
- February 24, 2005
- Terminated
- July 24, 2006
- Recall Number
- Z-0269-05
- Quantity
- 1150
- Firm Location
- Waukesha, WI
- Official Source
- View on FDA website ↗
Reason for Recall
The patient tables of the affected CT systems are supported by a hydraulic cylinder and a cylinder shaft (i.e. the upper shaft) which connects the tabletop to the table frame. GE received a report of a HiSpeed LX/I system table falling abruptly due to the filure of the upper shaft. No injuries were reported.
Lot / Code Info
Model or Catalog: 2297024, 2320315, 2244226, 2113694-2, 2115992-4, 2200290-2
Root Cause
Other
Action Taken
Inadequate materials used to manufacture the upper shaft of patient tables that can result in uncontrolled table motion.