RecallDepth

Roche brand OMNI 9 clinical chemistry analyzer; catalog numbers GD0475, GD0475R and GD047591.

This recall has been terminated (originally issued January 10, 2004).

Company
Roche Diagnostics Corp.
Recall Initiated
December 15, 2003
Posted
January 10, 2004
Terminated
August 2, 2004
Recall Number
Z-0328-04
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

If the sample volume is insufficient, the analyzer may issue erroneous results without any indication to the user that the problem is insufficient sample volume.

Distribution

United States.

Lot / Code Info

All units with software version 5.61 or below.

Root Cause

Other

Action Taken

Consignees were notified by letter mailed on or about December 15, 2003. Customers are instructed to assure an adequate sample volume until new software becomes available.