RecallDepth

Proplege Coronary Sinus Catheter model PR9, 11 French Introducer, kitted with PR9. Product Usage: The catheter introducer sheath kit provides access to the central venous circulation to facilitate catheter and guidewire insertion. The Proplege coronary sinus catheter provides retrograde cardioplegia during cardiopulmonary by-pass surgery.

This recall has been terminated (originally issued May 14, 2013).

Company
Edwards Lifesciences, Llc
Recall Initiated
January 25, 2013
Posted
May 14, 2013
Terminated
February 16, 2017
Recall Number
Z-1322-2013
Quantity
807
Firm Location
Draper, UT
Official Source
View on FDA website ↗

Reason for Recall

Edwards Lifesciences is recalling certain Proplege Coronary Sinus Catheter, model PR9 due to confirmed reports of blood leakage through the hemostasis valve in the introducer that is kitted with the catheter.

Distribution

Worldwide Distribution - USA Nationwide including the states of: MD, NY, PA, CA, FL, AZ, AL, TX, VA, OH, MO, NC, GA, TN, CO, OR, SC, ME, IL, HI, NJ, WA, WI, CT, KY, LA and Canada.

Lot / Code Info

Model PR9, lot numbers: 59385159, 59385164, 59385204, 59398000, 59398006, 59398008, 59400402, 59423199, 59423200, 59385159, 59385164, 59385204, 59398000, 59398006.

Root Cause

Device Design

Action Taken

Edwards Lifesciences sent an Urgent Field Safety Notice letter dated February 1, 2013. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review entire inventory for the affected product, quarantine and return to Edwards. Contact Customer Services at 800-424-3278 for any question, to obtain a Returned Goods Authorization number and replacement product.