cobas 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.

Company
Roche Molecular Systems, Inc.
Recall Initiated
April 1, 2013
Posted
May 21, 2013
Terminated
August 7, 2014
Recall Number
Z-1346-2013
Quantity
425 instruments
Firm Location
Branchburg, NJ

Reason for Recall

In rare instances, channel shifted results have been generated with cobas 4800 assays run on v1.1.1 of the cobas 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported

Distribution

Nationwide Distribution.

Lot / Code Info

Material Number 05200881001 - all lot/serial numbers.

Root Cause

Software design

Action Taken

Roche sent an "Urgent Medical Device Correction" (UMDC) notification with response form dated April 1, 2013, to all affected customers via UPS Ground on April 1, 2013. The letter identified the product, the problem, and the action to be taken by the customer. A contact number (Roche Diagnostic Technical Support) is provided for technical support 24 hours a day, seven days a week at phone number: 1-888-912-7090.

More recalls by Roche Molecular Systems, Inc.