ARTISTE and ONCOR linear accelerators with G31 mother board. ARTISTE and ONCOR are manufactured by Siemens Medical Solutions, Roentgenstrasse 19-21, Kemnath, Germany. ARTISTE, ONCOR and PRIMUS family of linear accelerators deliver X-ray photon and electron radiation for therapeutic treatment of cancer.

This recall has been terminated (originally issued January 11, 2012).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
November 9, 2011
Posted
January 11, 2012
Terminated
April 15, 2013
Recall Number
Z-0459-2012
Quantity
13 active devices
Firm Location
Concord, CA
Official Source
View on FDA website ↗

Reason for Recall

Reports of gantry rotation without receiving a movement command caused by an overload within the drive signal power line.

Distribution

Worldwide Distribution-- USA ( nationwide ) including the states of : MO, NC, OH, and OK and the countries of Turkey, Japan, Germany and Portugal

Lot / Code Info

Artiste MV system, part number 8139789 ONCOR Impression PLUS, part number 5857912

Root Cause

Component design/selection

Action Taken

Siemens Medical Solutions USA, Inc. sent a Customer Information letter dated November 9, 2011, by federal express. The letter identified the product the problem and the action needed to be taken by the customer. The HW Update included information regarding the potential problem and provide instructions in the Safety update. Further questions call (925) 602-8157.