Roche/Hitachi brand lactate reagent for the quantitative determination of L-lactate in plasma, cerebrospinal fluid or whole blood. Catalog # 1822837.

This recall has been terminated (originally issued October 15, 2003).

Company
Roche Diagnostics Corp.
Recall Initiated
September 2, 2003
Posted
October 15, 2003
Terminated
December 10, 2003
Recall Number
Z-0028-04
Quantity
21958
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Patient samples with very high lactate values may falsely be reported as within the normal range without the instrument giving a warning flag to the technician.

Distribution

United States, Canada and Germany

Lot / Code Info

All lots of reagent when used manually or on the Roche/Hitachi 902, 911, 912, 917 or Modular Analytics P module analyzers.

Root Cause

Other

Action Taken

Recall letters were sent to U.S. customers on 9/2/03 instructing them not to run lactate with these analyzers. An update notification letter was sent on or about 9/18/03 telling customers to discontinue use of the product for manual lactate determinations.