QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-4X3, Order Number G22654; used to dilate strictures of the biliary tree

Company
Wilson-Cook Medical Inc
Recall Initiated
November 3, 2023
Posted
December 1, 2023
Recall Number
Z-0459-2024
Quantity
553 devices
Firm Location
Winston Salem, NC

Reason for Recall

Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.

Distribution

US

Lot / Code Info

UDI/DI 00827002226548 (each), 10827002226545 (case), Lot Numbers: W4720848, W4721792, W4721795, W4722159, W4722981, W4723304, W4723846, W4724692, W4737576, W4738717, W4742525, W4743193, W4743605, W4743606, W4743607, W4743608, W4744022, W4744441, W4748043, W4748946, W4749748, W4750264, W4750265, W4750910, W4751772, W4751906, W4751914, W4753360, W4753361, W4753362, W4753365, W4753816, W4755138, W4756901, W4756903, W4756905, W4757732, W4758647, W4759511, W4759976, W4760715, W4761738, W4761740, W4761935, W4762150

Root Cause

Under Investigation by firm

Action Taken

Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice on 11/03/2023 by mail. The notice explained the problem with the device, risk to health, and requested the following actions be taken: Customer: Cease use, quarantine, and return all affected devices. If devices were further distributed, those to whom the devices were sent must be notified. Distributor: Quarantine all affected devices, notify your customers to whom you distributed devices. Return all devices to Cook Medical.