QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-4X3, Order Number G22654; used to dilate strictures of the biliary tree

This recall is currently active, issued December 1, 2023. It was issued by Wilson-Cook Medical Inc.

Company
Wilson-Cook Medical Inc
Recall Initiated
November 3, 2023
Posted
December 1, 2023
Recall Number
Z-0459-2024
Quantity
553 devices
Firm Location
Winston Salem, NC
Official Source
View on FDA website ↗

Reason for Recall

Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.

Distribution

US

Lot / Code Info

UDI/DI 00827002226548 (each), 10827002226545 (case), Lot Numbers: W4720848, W4721792, W4721795, W4722159, W4722981, W4723304, W4723846, W4724692, W4737576, W4738717, W4742525, W4743193, W4743605, W4743606, W4743607, W4743608, W4744022, W4744441, W4748043, W4748946, W4749748, W4750264, W4750265, W4750910, W4751772, W4751906, W4751914, W4753360, W4753361, W4753362, W4753365, W4753816, W4755138, W4756901, W4756903, W4756905, W4757732, W4758647, W4759511, W4759976, W4760715, W4761738, W4761740, W4761935, W4762150

Root Cause

Under Investigation by firm

Action Taken

Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice on 11/03/2023 by mail. The notice explained the problem with the device, risk to health, and requested the following actions be taken: Customer: Cease use, quarantine, and return all affected devices. If devices were further distributed, those to whom the devices were sent must be notified. Distributor: Quarantine all affected devices, notify your customers to whom you distributed devices. Return all devices to Cook Medical.