Product Name: Salivary Testosterone ELISA Model/Catalog Number: SLV-3013 Product Description: An enzyme immunoassay for the quantitative in vitro diagnostic measurement of free active testosterone in saliva.

Company
Drg International, Inc.
Recall Initiated
July 5, 2024
Posted
August 15, 2024
Recall Number
Z-2622-2024
Quantity
257 units
Firm Location
Springfield, NJ

Reason for Recall

Invalid Controls. The Low Control is out of range; too high which causes invalid runs that would result in repeating the runs and delays for reporting results to patients.

Distribution

US Nationwide distribution in the states of CA, IL, MN, PA, UT, VA.

Lot / Code Info

Model/Catalog Number: SLV-3013; UDI-DI: 04048474030136; Lot Nymbers: 66K014 and 66K123.

Root Cause

Nonconforming Material/Component

Action Taken

On July 8, 2024 Urgent Field Safety Notification letters were sent to customers. Action to be taken by Distributors and End-users: Please check your inventory for the respective lots 66K014 and 66K123. and stop using these kits immediately and promptly quarantine them. If you faced this issue of failed controls, kindly report with daya, and request a replacement for the kits used that were affected. Also, all unused kits still in your stock of the affected lot(s) will also be replaced. Immediately reply to email: qa@drg-international.com to confirm this notice and your compliance. Please indicate in the email the quantity of kits used which failed and quantity of stock remaining. We sincerely apologize for the inconvenience this may have caused you and/or your customer(s).