Total Thyroxine (T4) ELISA; REF EIA-1781

Company
Drg International, Inc.
Recall Initiated
October 18, 2021
Recall Number
Z-0310-2022
Quantity
314 total (270 US)
Firm Location
Springfield, NJ

Reason for Recall

Firm has identified two lots of product that do not meet product shelf life, resulting in failed Controls. The clinical impact may be a delay in diagnosis and a consumption of patient samples.

Distribution

Domestic: NY; International: TZ, VN, BD, CO, TT, YE, BO, DE, SV, GE, PY, TG, RU/UA, NI, DO, NG, EG & CM

Lot / Code Info

lot#: RN-60923 expiration date 12/31/2021; RN-60831 expiration date 12/31/2021

Root Cause

Nonconforming Material/Component

Action Taken

On 10/18/2021, DRG issued Medical Device Recall Notification letters and response forms to the End-user through confirmation of the Distributor. All customers were notified via e-mail. Actions Required: 1) Please immediately discontinue use or distribution of the affected lots. 2) If you have further distributed the affected product lots, please identify and immediately notify those customers at once of this product recall by providing a copy of the recall notice and response form. 3) If you have relabeled or rebranded the affected product lots, please follow your internal recall process and notify the appropriate regulatory bodies as required. 4) Disposal of all affected lots should be in accordance with applicable national, st ate, and local laws and regulations. 5) Complete and return the response form.