Abutment Screw Retrieval Instrument 3.0/NP - Product Usage: The instruments for abutment screw retrieval can be used if an abutment screw is broken and a part of the screw is left inside the implant screw channel. If the implant thread needs to be cleaned before placing a new screw, a screw tap repair can be used. The instruments are intended for both simple and advanced treatments.

Company
Nobel Biocare Usa Llc
Recall Initiated
March 14, 2019
Posted
February 19, 2020
Terminated
February 15, 2024
Recall Number
Z-1322-2020
Quantity
233 units
Firm Location
Yorba Linda, CA

Reason for Recall

Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.

Distribution

Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.

Lot / Code Info

(01)07332747084085(10)73585 (01)07332747084085(10)73430 Catalog # 37503

Root Cause

Packaging process control

Action Taken

On 3/27/2018 the firm issued a recall notice by email to affected consignees. The recall notification informed the following actions to be taken by the consignees: 1. Identify affected devices as detailed in the Customer Reply Form. 2. Do not use affected devices. 3. Send back the Customer Reply Form and the affected devices. The firm will replace affected devices free of charge. This letter needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. We kindly ask you to follow instructions provided in this Recall Letter and return the attached Customer Reply Form to the firm via email or fax at 714- 282-5073.