Situate Gauze RF and X-Ray detectable. Product Number: G0804-16P02CN-1 Situate premium cotton are single use cotton disposable that must be used with the Situate delivery system. The disposable have an attached RF Tag, which is an electrically passive inductor like device that is sewn into the green pouch that you will see on all Situate premium cotton. When activated by magnetic impulses from the scanning devices, the tag resonates a signal that allows for the detection of these cotton disposable.

This recall has been terminated (originally issued October 12, 2017).

Company
Covidien Llc
Recall Initiated
October 12, 2017
Terminated
May 12, 2020
Recall Number
Z-1130-2018
Quantity
20,800
Firm Location
North Haven, CT
Official Source
View on FDA website ↗

Reason for Recall

This voluntary recall is being conducted due to the potential for failure to detect the affected Situate" gauze - RF and Xray detectable products by the Situate" detection console due to a manufacturing error.

Distribution

U.S.

Lot / Code Info

Lot Number: 170704A UDI: 20884521700861

Root Cause

Nonconforming Material/Component

Action Taken

All consignees were notified via FedEx or certified mail on October 12, 2017 and the letter 'Urgent Field Corrective action Notice' informs customers of the potential for failure to detect the affected Situate" gauze RF and X-Ray detectable products by the Situate" detection console and the actions they should take. Customers are requested to acknowledge receipt and understanding of the Urgent Field Correction Action Notice. For further questions, please call (203) 492-5000.