Toshiba Ultimax DREX-ULT80 X-ray generator

This recall has been terminated (originally issued May 19, 2016).

Company
Toshiba American Medical Systems Inc
Recall Initiated
May 19, 2016
Terminated
June 14, 2018
Recall Number
Z-1398-2017
Quantity
254
Firm Location
Tustin, CA
Official Source
View on FDA website ↗

Reason for Recall

It was discovered that the generator of the system could possibly terminate the exposure prematurely during an examination

Distribution

USA ( nationwide ) Distribution to the states of : WA, OR, CA, OH, IL, CT, IA, KY, AL, LA, HI, PA, NJ, DE, MN, DC, NY, MI, SC, MD, TN, OK, WI, UT, ID, NV, TX, AL, IN, WV, KS, MO, CO, MT, MS, NC, GA, NH,UT, FL, AZ and ND.

Lot / Code Info

Model : KXO-80XM

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Toshiba America Medical Systems planned action to bring defect into compliance : 1. A notification letter will be sent to customers with affected systems. This letter includes the statement that the repairs will be made at no cost to the customer. 2. Modified firmware will be installed on the systems to prevent the defect from recurring. 3. The consignees will be contacted by a Toshiba service representative for an appointment when the modified firmware becomes available 4. The consignee will be informed that if this problem occurs before the new firmware is installed; they should reset the x-ray generator to complete the exam; and 5. The CAP will be completed by May 1, 2017. For further questions please call (714) 730-5000