Neodent Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 -intended for dental implant Item 138.089
- Company
- Straumann Usa Llc
- Recall Initiated
- November 30, 2021
- Terminated
- February 5, 2025
- Recall Number
- Z-0512-2022
- Quantity
- 7 pieces
- Firm Location
- Andover, MA
Reason for Recall
Mislabeled: Carton box item 138.089- Easypack Helix GM Acqua 3.5X8/Smart 4.5X2.5 has an implant with measure of 3.5x13mm; and Carton Box (138.107- Easypack Helix GM Acqua has a 3.5x8mm implant instead of a 3.5x 13mm implant
Distribution
US Nationwide Distribution in the states of FL, IL, OH
Lot / Code Info
Lot GNV17 UDI: 7899878052722
Root Cause
Packaging change control
Action Taken
Neodent issued Medical Device Recall letter initial contact by phone initiated on November 30th, 2021; followed by Confirmation written letter via UPS December 3rd 2021. Letter states reason for recall, health risk and action to take: Search in your stock for units according to the description on the table above. " If the implant has not yet been used, identify and segregate the affected units, open or closed, in your stock. Return the product using the label included in this mailing, refer to Field Action . Upon receipt of the product by Straumann, replacement product will be sent. If the product is not found in your inventory, indicate on the Form below. " If the product has been implanted and no issue was identified, it is not necessary to remove, additional patient follow-up is not required. Indicate this on the Customer Confirmation Form (included) " If the product has been installed and an issue has been reported by the patient, for example: numbness, pain or any other symptom, please communicate immediately to Straumann. " For either cases, complete and return the enclosed Customer Confirmation Form to Action 003/2021. 3.1. DATA FOR PRODUCT RETURN: Straumann USA Regulatory Affairs Action 003/2021 Product Return 60 Minuteman Road, Andover, MA 01810