GM Helix Acqua Implant, Article Number: 140.985

Company
Straumann Usa Llc
Recall Initiated
May 1, 2025
Posted
May 28, 2025
Recall Number
Z-1863-2025
Quantity
207 units (54 US, 153 OUS)
Firm Location
Andover, MA

Reason for Recall

It is possible that a package labelled as an 11.5 mm implant may contain a 13 mm implant instead.

Distribution

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, Il KS, LA, MA, MD, MI, NE, NH, NJ, NY, OH, OK, RI, SC, TN, TX, UT, WI and the countries of Australia, Croatia, France, Hungary, India, Italy, Lebanon, Lithuania, New Zealand, Portugal, Qatar, Serbia, Spain, Tunisia, UAE, United Kingdom, Vietnam.

Lot / Code Info

Article Number: 140.985; UDI-DI: (01)7899878024569(11)220125(17)270124(10)JPZ75; Lot number: JPZ75;

Root Cause

Under Investigation by firm

Action Taken

On May 1, 2025 URGENT FIELD SAFETY ACTION letters were sent to customers. Action to be taken: 1. If article 140.984 Lot JPZ21 or 109.985 Lot JPZ75 are still in your inventory, then stop use/distribution of the product immediately and quarantine/segregate physically. 2. If the product was used and no complications or problems were identified, of the patient did not present symptoms there is no need for any additional action with the patient. 3. If article 140.984 Lot JPZ21 or 109.985 Lot JPZ75 are still in your inventory, return it for a credit or replacement as indicated on the Customer Confirmation Form. 4. In the product is not found in your stock, indicate this on the enclosed Customer Confirmation Form. 5 In all cases, complete and return the enclosed Customer Confirmation Form with the affected product using the return UPS label provided to: Straumann USA, LLC; 60 Minuteman Road, Andover, MA 01810. Transmission of the Field Safety Notice: This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the affected devices have been transferred. The FDA has been informed about this Field Safety Corrective Action. We apologize for any inconvenience that this may cause.

More recalls by Straumann Usa Llc

View all recalls by this company →