Puritan Bennett 980 Series Ventilator, 980X2ESDIEC

Company
Covidien Llc
Recall Initiated
November 4, 2021
Posted
December 9, 2021
Terminated
July 9, 2024
Recall Number
Z-0351-2022
Quantity
278 total
Firm Location
Mansfield, MA

Reason for Recall

Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.

Distribution

Worldwide distribution - US Nationwide.

Lot / Code Info

GTIN/UDI - 10884521174177 Serial Numbers: 35B1700627 35B1700628 35B1700629 35B1700630 35B1700631 35B1700632

Root Cause

Reprocessing Controls

Action Taken

On November 4, 2021, the firm began to notify customers through Urgent Medical Device Correction letters. Customers were informed of the device defect that may lead to the ventilator becoming inoperable during use. Actions you should take Immediately discontinue use of the affected ventilators. Remove the affected ventilators from clinical service and quarantine the affected ventilators until such time that a Medtronic Technical Service Engineer inspects and replaces the affected printed circuit board assemblies. Notify all personnel in all care environments in which the affected PB980 series ventilators are used about this medical device correction. If your facility has distributed affected PB980 series ventilators to other persons or facilities, please promptly forward a copy of this letter to those recipients. Complete the attached form and return it as directed to confirm your receipt and understanding of this information. If you are aware of any incidents related to this issue, please contact our Technical Support Department immediately at 1-800-255-6774, option 4, and then option 1 to provide information regarding those events so regulatory reporting obligations can be fulfilled. Work with the Medtronic Technical Support Department if you require assistance finding alternative ventilation devices. Beginning November 5, 2021, Medtronic Technical Service will contact all affected customers to schedule a service appointment for all potentially affected ventilators. If you have any questions regarding this communication, please contact your Medtronic representative or Technical Service at 1-800-255-6774.

More recalls by Covidien Llc

View all recalls by this company →