Sterile Procedural Trays, labeled as the following: a. MAJOR CDS-LF b. BURN PACK-LF c. BURN PACK d. SMALL BURN
This recall is currently active, issued February 6, 2023. It was issued by Medline Industries, Lp - Northfield.
- Recall Initiated
- November 17, 2022
- Posted
- February 6, 2023
- Recall Number
- Z-1083-2023
- Quantity
- 100 cases (156 units)
- Firm Location
- Northfield, IL
- Official Source
- View on FDA website ↗
Reason for Recall
Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.
Lot / Code Info
a. MAJOR CDS-LF, Item Number: CDS983861D, Case UDI/GTIN: 40193489524353, Unit UDI/GTIN: 10193489524352, Lot Number: 22JDB004; b. BURN PACK-LF, Item Number: DYNJ0781678S, Case UDI/GTIN: 40193489226349, Unit UDI/GTIN: 10193489226348, Lot Number: 22IMC056; c. BURN PACK, Item Number: DYNJ82534, Case UDI/GTIN: 40195327198504, Unit UDI/GTIN: 10195327198503, Lot Number: 22HBI219; d. SMALL BURN, Item Number: DYNJ901793L, Case UDI/GTIN: 40193489955065, Unit UDI/GTIN: 10193489955064, Lot Number: 22KBB743,22IBH124
Root Cause
Device Design
Action Taken
Medline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.