Merge OrthoPACS software.

This recall has been terminated (originally issued May 24, 2017).

Company
Merge Healthcare, Inc.
Recall Initiated
May 24, 2017
Terminated
June 28, 2019
Recall Number
Z-2620-2017
Quantity
98 sites potentially have the affected versions
Firm Location
Hartland, WI
Official Source
View on FDA website ↗

Reason for Recall

A migrated study that has annotations will display the annotations, but when a prior study is loaded that should have annotations, the second study will not display annotations.

Distribution

Distribution was made nationwide to medical facilities. There was no foreign/government/military distribution.

Lot / Code Info

Versions 6.6, 6.6.1, and 6.6.2

Root Cause

Software design

Action Taken

The recalling firm issued a letter dated 5/19/2017 via email on 5/24/2017 or via certified mail if they did not have an email address for the customer.