Medfusion Syringe Pump Model 3500, Software Version V6.0. Infusion pump.
- Company
- Smiths Medical Asd Inc.
- Recall Initiated
- June 26, 2020
- Posted
- August 6, 2020
- Recall Number
- Z-2735-2020
- Quantity
- 17019 units
- Firm Location
- Minneapolis, MN
Reason for Recall
Inaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.
Distribution
Worldwide distribution - United Arab Emirates, United States, Afghanistan, Albania, Australia, Belgium, Bermuda, Canada, China, Germany, Denmark, Great Britain, Hong Kong, Indonesia, Ireland, India, Korea, Kuwait, Lebanon, Mexico, Malaysia, New Zealand, Oman, Panama, Philippines, Pakistan, Poland, Saudi Arabia, Trinidad and Tobago
Lot / Code Info
Model Numbers: 3500-0600-00, 3500-0600-01, 3500-0600-249, 3500-0600-50, 3500-0600-51, 3500-0600-82. Software Version V6.0
Root Cause
Software design
Action Taken
Consignees are being contacted via direct mailing and email. Consignees are asked to review and complete the Recall Response form indicating understanding of notification. Distributors are asked to forward the notification to parties to whom product has been distributed. Smiths Medical will send out 2 additional notifications by email. Phone calls will be made to non-responders.