RecallDepth

Multifire Endo Hernia 0 12mm Stapler with 4.0mm Disposable Loading Unit , Part No. 174027 - Product Usage: has application in endoscopic general surgery procedures for fixation of mesh in the repair of hernial defects and in other surgical specialties for approximation of tissue(s).

Company
Covidien Llc
Recall Initiated
May 12, 2020
Recall Number
Z-2255-2020
Quantity
12,610 total
Firm Location
Mansfield, MA

Reason for Recall

The product is recalled due to a manufacturing assembly error that may cause the staples to fail to properly advance within the staple cartridge when the instrument handle is compressed.

Distribution

Worldwide distribution.

Lot / Code Info

Lots P9J0368Y  P9J0369Y  P9J0517Y  P9K1300Y

Root Cause

Process control

Action Taken

All consignees were notified by letter titled "Urgent Medical Device Recall" delivered via Federal Express and/or certified mail beginning May 12, 2020. The letter informs customers of the recall of specific item codes and production lots of Multifire hernia staplers due to a manufacturing assembly error. Use of a device with this condition may result in the failure of staples to properly advance within the staple cartridge when the instrument handle is compressed. Customers are instructed to take the following actions: " Immediately quarantine and discontinue use of the affected item code and lot. " Return affected product. " Promptly forward the information from the customer communication to any recipients to whom they have distributed any of the affected product. " Complete the Recalled Product Return Form even if you do not have inventory. If you have any questions about this event, you may contact Medtronic via email at rs.assurancequality@medtronic.com

More recalls by Covidien Llc

View all recalls by this company →