RecallDepth

EC-5000 Excimer Laser Corneal Surgery System (EC-5000, EC 5000 CXIII, EC-5000 Quest. Opthalmic: NIDEK EXCIMER LASER CORNEAL SURGERY SYSTEM, model EC-5000, is an ophthalmic laser system used for correction of corneal refraction and ablation of the corneal surface.

This recall has been terminated (originally issued June 4, 2015).

Company
Nidek Inc
Recall Initiated
March 24, 2015
Posted
June 4, 2015
Terminated
April 5, 2016
Recall Number
Z-1710-2015
Quantity
114 units
Firm Location
Fremont, CA
Official Source
View on FDA website ↗

Reason for Recall

Multi-stage treatment option for the EC-5000 Operator's Manual was not approved by the FDA.

Distribution

Us distribution only.

Lot / Code Info

Alll devices currently in use by consignees.

Root Cause

No Marketing Application

Action Taken

Nidek sent an Device Correction letter dated April 30th, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Nidek will contact customers to schedule an appointment to replace the operators manual and implementing a software lockout device on the device so that the multi stage feature is no longer available. If a Nidek rep does not contact you before April 30, 2015, please call 800-223-9044 to schedule an appointment, All non responding customers.will be contacted three time via Fed Ex letter